Planer CT37 STAX Installation Guide
Status Monitoring:
Step 1: Equipment needed
Step 2: Find Planer CT37 STAX external alarm port
Step 3: Link Planer CT37 STAX to ViGIE
Step 4: Configure an alarm and perform a test
Step 1: Equipment needed
- Airo Sensor 202041
- Planer CT37 STAX Equipment Status Cable/Adapter
Step 2: Find Planer CT37 STAX external alarm port
The Alarm Port is located at the back of the unit.
Step 3: Link Planer CT37 STAX to ViGIE
Connect the Planer CT37 STAX and the External Status Adapter to the Airo Sensor 202041 External adapter, using the NO (Normally Open) contacts.
Then the Adapter is ready to connect to the Analog Alarm Output from Planer CT37 STAX .
Step 4: Configure an alarm and perform a test
1st - Find your sensor using the ViGIE Platform and configure an alarm profile.
2nd - Check if the Planer CT37 STAX status match the ViGIE platform
Planer CT37 STAX OK = ViGIE OK
If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap
My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.
3rd - Simulate an event
After simulate an event, please ensure that a new ViGIE Alarm was created/sent.
Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.
Whenever the equipment alarms, the external alarm output will be activated.