Astec CUBE AD-3100 Installation Guide
Status Monitoring:
Step 1: Equipment needed
Step 2: Find Astec CUBE AD-3100 external alarm port
Step 3: Link Astec CUBE AD-3100 to ViGIE
Step 4: Configure an alarm and perform a test
Step 1: Equipment needed
- Airo Sensor 202041
- Astec CUBE AD-3100 Equipment Status Cable/Adapter
Step 2: Find CUBE AD-3100 external alarm port
The Alarm Ports are located at the back of the unit.
Step 3: Link Astec CUBE AD-3100 to ViGIE
Connect the Astec CUBE AD-3100 and the External Status Adapter to the Airo Sensor 202041 External adapter, using the NO (Normally Open) contacts.
Then the Adapter is ready to connect to the Analog Alarm Output from Astec CUBE AD-3100 .
Note: There are many signal outputs (1 per chamber) and a Power Monitor Output.
Step 4: Configure an alarm and perform a test
1st - Find your sensor using the ViGIE Platform and configure an alarm profile.
2nd - Check if the Astec CUBE AD-3100 status match the ViGIE platform
Astec CUBE AD-3100 = ViGIE OK
If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap
My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.
3rd - Simulate an event
After simulate an event, please ensure that a new ViGIE Alarm was created/sent.
Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.
Whenever the equipment alarms, the external alarm output will be activated.
Analog Monitoring:
Find the Analog Outputs and link the connectors.
Note: 1 Analog Output per chamber