Benchtop Incubators - Genea Biomedx Geri



Genea Biomedx Geri Installation Guide


For Status Monitoring:


  • Step 1: Equipment needed

  • Step 2: Find Geri external alarm port

  • Step 3: Link Geri to ViGIE

  • Step 4: Configure an alarm and perform a test



Step 1: Equipment needed 


-    Airo Sensor 202041

-    External Status Adapter 



Step 2: Find Geri External Alarm Contact


The External Alarm Contact is located at the back of the unit.





Step 3: Link Geri to ViGIE 


Connect the NO (Normally Open) Geri Alarm Contact to the External Status Adapter to the Airo Sensor 202041


Additional Info: Please use the NO (Normally Open) Alarm Contact.



Step 4: Configure an alarm and perform a test 


1st - Find your sensor using the ViGIE Platform and configure an alarm profile.

2nd - Check if the Geri status match the ViGIE platform


Geri OK = ViGIE OK


If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap

My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.



3rd - Simulate an event


After simulate an event, please ensure that a new ViGIE Alarm was created/sent. 

Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.


This device will alarm if power has not been restored in approximately 100 seconds.



For External Probe Monitoring:


Find the external temperature probe port then insert the probe.


Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article