Benchtop Incubators - Astec EC6S-MD



Astec EC6S-MD Installation Guide



  • Step 1: Equipment needed

  • Step 2: Find EC6S-MD external alarm port

  • Step 3: Link EC6S-MD to ViGIE

  • Step 4: Configure an alarm and perform a test



Step 1: Equipment needed 


-    Airo Sensor 202041

-    Astec EC6S-MD Equipment Status Cable/Adapter



Step 2: Find EC6S-MD external alarm port 


The Alarm Port is located at the back of the unit.




Step 3: Link EC6S-MD to ViGIE 


Connect the  EC6S-MD and the External Status Adapter to the Airo Sensor 202041 External adapter, using the NO (Normally Open) contacts.

Then the Adapter is ready to connect to the Analog Alarm Output from EC6S-MD.


Step 4: Configure an alarm and perform a test 


1st - Find your sensor using the ViGIE Platform and configure an alarm profile.

2nd - Check if the EC6S-MD status match the ViGIE platform


Esco Miri OK = ViGIE OK


If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap

My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.



3rd - Simulate an event


After simulate an event, please ensure that a new ViGIE Alarm was created/sent. 

Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.


Whenever the equipment alarms, the external alarm output will be activated.