Benchtop Incubators - Cook Medical MINC Mini Incubator



Cook Medical MINC Installation Guide



  • Step 1: Equipment needed

  • Step 2: Find MINC external alarm port

  • Step 3: Link MINC to ViGIE

  • Step 4: Configure an alarm and perform a test



Step 1: Equipment needed 


-    Airo Sensor 202041

-    External Status Adapter 



Step 2: Find MINC external alarm port 


The Alarm Port is located at the back of the unit.




Step 3: Link MINC to ViGIE 


1st - Connect the 2 Wire to the 2 Pin Phoenix Connector


2nd - Connect the  External Status Adapter to the Airo Sensor 202041




Step 4: Configure an alarm and perform a test 


1st - Find your sensor using the ViGIE Platform and configure an alarm profile.

2nd - Check if the MINC status match the ViGIE platform


MINC OK = ViGIE OK


If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap

My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.



3rd - Simulate an event


After simulate an event, please ensure that a new ViGIE Alarm was created/sent. 

Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.


This device will alarm if : 
- Low inlet CO2 Pressure (15 minutes later)
- No gas flow (5 minutes later)
- Temperature (2 minutes later)
- Power failure (2 minutes later)