Benchtop Incubators - K-Systems G185



K-Systems G185 Installation Guide



  • Step 1: Equipment needed

  • Step 2: Find G185 external alarm port

  • Step 3: Link G185 to ViGIE

  • Step 4: Configure an alarm and perform a test



Step 1: Equipment needed 


-    Airo Sensor 202041

-    External Status Adapter (pre-assembled with 3.5mm Jack)



Step 2: Find G185 external alarm port 


The Alarm Port is located at the back of the unit.




Step 3: Link G185 to ViGIE 


Connect the 3.5mm Jack to G185 and the External Status Adapter to the Airo Sensor 202041


Additional Info: The output is NC (Normally Closed) and non configurable.



Step 4: Configure an alarm and perform a test 


1st - Find your sensor using the ViGIE Platform and configure an alarm profile.

2nd - Check if the G185 status match the ViGIE platform


G185 OK = ViGIE OK


If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap

My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.



3rd - Simulate an event


After simulate an event, please ensure that a new ViGIE Alarm was created/sent. 

Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.


This device will alarm if : 
- An alarm is activated 
- Incubator is powered off