Benchtop Incubators - Astec EC9 Installation Guide


Astec EC9 Installation Guide



Status Monitoring:


  • Step 1: Equipment needed

  • Step 2: Find Astec EC9 external alarm port

  • Step 3: Link Astec EC9 to ViGIE

  • Step 4: Configure an alarm and perform a test



Step 1: Equipment needed 


-    Airo Sensor 202041

-    Astec EC9 Equipment Status Cable/Adapter



Step 2: Find EC9 external alarm port 


The Alarm Port is located at the back of the unit.





Step 3: Link Astec EC9 to ViGIE 


Connect the Astec EC9 and the External Status Adapter to the Airo Sensor 202041 External adapter, using the NO (Normally Open) contacts.



Step 4: Configure an alarm and perform a test 


1st - Find your sensor using the ViGIE Platform and configure an alarm profile.

2nd - Check if the Astec EC6 status match the ViGIE platform


Astec EC9  = ViGIE OK


If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap

My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.



3rd - Simulate an event


After simulate an event, please ensure that a new ViGIE Alarm was created/sent. 

Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.


Whenever the equipment alarms, the external alarm output will be activated



Analog Monitoring:


Find the 18 Pin Port and link the connector.



Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article