Benchtop Incubators - Astec CCM-iBIS


Astec CCM-iBIS Installation Guide



Status Monitoring:


  • Step 1: Equipment needed

  • Step 2: Find Astec CCM-iBIS external alarm port

  • Step 3: Link Astec CCM-iBIS to ViGIE

  • Step 4: Configure an alarm and perform a test



Step 1: Equipment needed 


-    Airo Sensor 202041

-    Astec CCM-iBIS Equipment Status Cable/Adapter



Step 2: Find CCM-iBIS external alarm port 


The Alarm Port is located at the back of the unit.





Step 3: Link Astec CCM-iBIS to ViGIE 


Connect the Astec CCM-iBIS and the External Status Adapter to the Airo Sensor 202041 External adapter, using the NO (Normally Open) contacts.

Then the Adapter is ready to connect to the Analog Alarm Output from Astec CCM-iBIS .


Step 4: Configure an alarm and perform a test 


1st - Find your sensor using the ViGIE Platform and configure an alarm profile.

2nd - Check if the Astec CCM-iBIS status match the ViGIE platform


Astec CCM-iBIS  = ViGIE OK


If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap

My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.



3rd - Simulate an event


After simulate an event, please ensure that a new ViGIE Alarm was created/sent. 

Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.


Whenever the equipment alarms, the external alarm output will be activated.


Analog Monitoring:


Find the 18 Pin Port and link the connector.

 