Esco Miri TL Installation Guide
Step 1: Equipment needed
Step 2: Find Esco Miri TL external alarm port
Step 3: Link Esco Miri TL to ViGIE
Step 4: Configure an alarm and perform a test
Step 1: Equipment needed
- Airo Sensor 202041
- External Status Adapter (pre-assembled with 3.5mm Jack)
Step 2: Find Esco Miri TL external alarm port
The Alarm Port is located at the back of the unit.
Step 3: Link Esco Miri TL to ViGIE
Connect the 3.5mm Jack to Esco Miri TL and the External Status Adapter to the Airo Sensor 202041
Additional Info: The output is NC (Normally Closed) and non configurable.
Step 4: Configure an alarm and perform a test
1st - Find your sensor using the ViGIE Platform and configure an alarm profile.
2nd - Check if the Esco Miri TL status match the ViGIE platform
Esco Miri OK = ViGIE OK
If the status is mismatched (eg: Equipment Alarm but ViGIE OK) please raise a ticket with ViGIE Helpdesk Team
.
To: helpdesk@vigiesolutions.com
Subject: Status Sensor Label Swap
My Airo Sensor 202041 with Serial Number: (please fill in) needs a Label Swap.
3rd - Simulate an event
After simulate an event, please ensure that a new ViGIE Alarm was created/sent.
Then a recovery should be tested as well, performing a status follow-up using ViGIE Platform.
This device will alarm if :
- Temperature alarm
- Gas alarm for CO2 or O2 levels
- Low-pressure or high-pressure alarm for CO2 and N2 gases
- Power supply to the unit is suddenly cut